- Continuous saturation pressure measurement capability.
- Oil-free high vacuum system for analysis.
- Separate oil-free high vacuum system for degassing.
- Capability for analysis gases such as N₂, Ar, Kr, CH₄, CO₂, H₂, CO, NH₃ and other non-corrosive gases.
- A single transducer is used at a time; the same transducer measures both the dosed gas pressure and the equilibrated pressure.
- Dedicated pressure transducer for continuous P₀ monitoring.
- Dead-end and through-pore analysis capability.
BET-201A-2OS
PMI’s BET Sorptometer has a wide range of applications across multiple industries worldwide.Applicable industries include Catalyst and Carbon, Oil & Gas/Geoscience, Battery and Fuel Cells, Chemical, Ceramic, Filtration, Paper, Pharmaceutical/Medical, Powder Metallurgy, and Automotive industries.
Specifications:
| Isotherm | Up to unlimited data points (per station), adsorption and desorption,hysteresis plot |
| Gas Inlet Ports | 5 |
| Pressure Measurement Ranges & Resolutions | Low: 0 to 0.1 mmHg transducer (typical): 0.0000001 mmHg resolution
Medium: 0 to 10 mmHg transducer (typical): 0.00001 mmHg resolution High: 0 to 1000 mmHg transducer (typical): 0.001mmHg resolution |
| Pore Size Range | 3.5 Å – 500 nm |
| Surface Area Range (in m²/g) | 0.0005 (Krypton) to no known upper limit |
| Surface Area Range (in m²/g) | 0.01 (Nitrogen) to no known upper limit |
| Micropore Volume (min) | 0.0001 cc/g |
| Adsorption Temperature | −195.6 °C (Liquid Nitrogen) |
| Power Requirements | 110/220 VAC, 50/60 Hz |